# Phantom 4 Multispectral

## Before Starting

{% hint style="info" %}
Reflectance panel images should be captured immediately before and/or after drone survey data collection.
{% endhint %}

***

## Instructions

1. Power on the drone and remote controller.&#x20;
2. Place the reflectance panel in an open and unobstructed area.
3. Connect your iPad to the remote controller and open the DJI GS Pro app on the iPad.
4. In the DJI GS Pro app Press the "My Missions" button.

<div align="center"><figure><img src="/files/IIXC31RGNlwhwdWOLCdo" alt="" width="563"><figcaption></figcaption></figure></div>

5. Press the drone icon

<figure><img src="/files/3RStFwpPPSjLdqCn1B8Q" alt="" width="563"><figcaption></figcaption></figure>

6. Press the camera view.

<figure><img src="/files/p8NhVtiERjIk6WaBtnKs" alt="" width="563"><figcaption></figcaption></figure>

7. Press the three line icon to open the image settings panel. Ensure that the camera is set to Auto.

<figure><img src="/files/qtOlO2f8ENkDo2ZVoqgH" alt="" width="563"><figcaption><p>Image Settings Panel</p></figcaption></figure>

8. Pick up aircraft and hold at waist height over calibration panel.
   * **Ensure aircraft is not shadowing the reflectance panel.**
9. Press the image capture button on the remote controller or in the DJI GS Pro app to take the calibration photo.
10. Set down the aircraft and pick up the panel.

## Modification of Calibration Image Name before upload into FieldAgent

<img src="https://support.senterasensors.com/home/~gitbook/image?url=https%3A%2F%2Fimg.zohostatic.com%2Fzde%2Fstatic%2Fimages%2Fcaution.png&#x26;width=40&#x26;dpr=3&#x26;quality=100&#x26;sign=41289ee0&#x26;sv=2" alt="" data-size="line"> <mark style="background-color:$danger;">This is only for if you are importing the calibration photo's along with your data into FieldAgent</mark>

In order for FieldAgent to realize there are calibration photo's from DJI sensor when uploading the following has to be done.

1. Click on File Explorer
2. Rename the files to have "CAL\_XXXX", see example below.

<div align="left"><figure><img src="/files/yUO2gem29YwOxuHbLtFx" alt=""><figcaption></figcaption></figure></div>

3. Navigate to each band and modify the calibration photo's.
4. Verify that calibration panel photo's were imported, i.e., When importing it should show in FieldAgent Desktop under survey details how many calibration panel photo's were imported. You will need to click on the arrow by where it says the how many images in the survey, it will then show the survey details.

<div align="left"><figure><img src="/files/eKlFkqjDvQ112It21JE8" alt=""><figcaption></figcaption></figure></div>

<div align="left"><figure><img src="/files/FWbKoOEULdUrsGbG5SkL" alt=""><figcaption></figcaption></figure></div>

## Importing data into Pix4D

If you are importing the calibration photo's into Pix4D you will need to add the following coefficients in order to correct the images, see below.

![](https://desk.zoho.com/api/v1/threads/280672000112877129/inlineImages/edbsn5798fce9230b029b45607ad9fc00b7c4081e2da33e0321f1691467d00efff20be31cb643758b4c7c2c02e883fcd625acdcb4a6e8c7c36126e6039dd5c605f5a0196c22038f17b6da46965f6692103e24?et=19de3afab88\&ha=d8015d07b32c06dfa9f32e6f6ae1bfc9623c832186cc4870617554cb92e16248\&f=1.png)\
\
Older Panel:\
NIR (840nm): 0.10561862\
Red (650nm): 0.10533071\
RE (730nm): 0.10511254\
Green (560nm): 0.10541804\
\
![](https://desk.zoho.com/api/v1/threads/280672000112877129/inlineImages/edbsn5798fce9230b029b45607ad9fc00b7c4081e2da33e0321f1691467d00efff20be31cb643758b4c7c2c02e883fcd625acca0a3d5e51d01165b477ba49a3b42488db42b4ab6588222518e25fc8ae8d5474?et=19de3afab89\&ha=b7f1e25963ade01361d6cc61a0f341cefb51a850053b44ec762c39b4e7d65ceb\&f=2.png)\
\
Newer Panel:\
NIR (840nm): 0.1050492\
Red (650nm): 0.10819064\
RE (730nm): 0.10690784\
Green (560nm): 0.1096772

Then follow this guide in order to add the coefficients to the bands. <https://support.pix4d.com/hc/en-us/articles/203891879#label3> .


---

# Agent Instructions: Querying This Documentation

If you need additional information that is not directly available in this page, you can query the documentation dynamically by asking a question.

Perform an HTTP GET request on the current page URL with the `ask` query parameter:

```
GET https://support.senterasensors.com/home/sentera-calibrated-reflectance-panel/sentera-calibrated-reflectance-panel/phantom-4-multispectral.md?ask=<question>
```

The question should be specific, self-contained, and written in natural language.
The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.